Determination of blood sirolimus concentrations in liver and kidney transplant recipients using the Innofluor® fluorescence polarization immunoassay: Comparison with the microparticle enzyme immunoassay and high-performance liquid chromatography-ultraviol

  • LORENA BOUZAS Unidad Monitorización Fármacos, Laboratorio Central, Hospital Clínico Universitario, Santiago de Compostela, Spain
  • JESÚS HERMIDA Unidad Monitorización Fármacos, Laboratorio Central, Hospital Clínico Universitario, Santiago de Compostela, Spain
  • J. CARLOS TUTOR Unidad Monitorización Fármacos, Laboratorio Central, Hospital Clínico Universitario, Santiago de Compostela, Spain
Keywords: Fluorescence polarization immunoassay, high-performance liquid chromatography, microparticle enzyme immunoassay, sirolimus

Abstract

Background. Although high-performance liquid chromatography (HPLC) is the method of choice for blood sirolimus determination, the microparticle enzyme immunoassay (MEIA) run on the IMx® analyser is widely used in therapeutic monitoring of this immunosuppressant agent. The aim of our study was to evaluate the possible determination of sirolimus using the fluorescence polarization immunoassay (FPIA) commercialized for everolimus quantification.

Methods. Sirolimus concentrations were determined in whole-blood samples from liver and kidney transplant recipients using the Innofluor® Certican® FPIA (Seradyn Inc.) run on a TDx® analyser (Abbott Laboratories), Sirolimus MEIA run on an IMx® analyser (Abbott Laboratories), and HPLC (UV detection) methods.

Results. The Innofluor® FPIA has a similar cross-reactivity with everolimus and sirolimus, and the within- and between-run coefficients of variation obtained for sirolimus determination were 2.7%–13.3%. In analysing different blood samples from liver and kidney transplant patients the linear regressions obtained were: FPIA = 1.12 HPLC + 0.43 (n=104, r=0.874), MEIA = 1.14 HPLC (n=146, r=0.892), and FPIA = 1.00 MEIA + 0.29 (n=106, r=0.941). Better correlation coefficients were obtained between the methods in the liver transplant samples (r≥0.900) than in the kidney transplant samples (r≥0.849). No significant effect was found for sirolimus clearance or the blood hematocrit on the relationship between the results produced by both immunoassays and HPLC.

Conclusion. The Innofluor® FPIA is a valid alternative with an analogous performance to the MEIA for the therapeutic monitoring of sirolimus.

Downloads

Download data is not yet available.
Published
2009-07-12
How to Cite
BOUZAS L., HERMIDA J., & TUTOR J. C. (2009). Determination of blood sirolimus concentrations in liver and kidney transplant recipients using the Innofluor® fluorescence polarization immunoassay: Comparison with the microparticle enzyme immunoassay and high-performance liquid chromatography-ultraviol. Upsala Journal of Medical Sciences, 114(1), 55-61. https://doi.org/10.1080/03009730802608254