6.1.2.1 Clinical Goals for Measurement of Hemoglobin in Primary Care as Assessed by Paper Vignettes

  • Geir Thue
  • Sverre Sandberg

Abstract

A random sample of 10% (273) Norwegian general practitioners (GP) received a questionnaire including 11 case stories. For each case the GP was asked to fill in an action value for hemoglobin which should represent the minimal change considered necessary by that general practitioner to take action towards the patient. These minimal changes could then be used for calculation of imprecision goals assuming that there was no change in the bias of the two measurements. The analytical imprecision goals (CV) derived from the different case stories varied between 0% and 7.8% with a mean of 2.8% accepting a false positive rate of 5%. The GPs assessment of a single case story also varied substantially.

Downloads

Download data is not yet available.
Published
1993-09-01
How to Cite
Thue G., & Sandberg S. (1993). 6.1.2.1 Clinical Goals for Measurement of Hemoglobin in Primary Care as Assessed by Paper Vignettes. Upsala Journal of Medical Sciences, 98(3), 331-334. https://doi.org/10.3109/03009739309179328
Section
Original Articles