6.1.3.3 Quality Specifications for pH and Blood Gases

  • Johan Kofstad

Abstract

This presentation is built upon experience with use of blood gas results in clinical work during 25 yeaqexperience from the external quality control of 120 blood-gas-analyzers participating in the Norwegian Control Survey 1984–1991, and at last 3 different pH/blood gas inter-analyzer comparisons in Oslo and Bergen in 1980, 1990 and 1992 (1, 2, 3). It is discussed the pure biological variations of the three parameters pH, pC0, and PO, as well as the variations in results caused by preanalytical and analytical factors. My proposal for goal of quality for determinations of blood gases (the pure analytical variations, given as EA, Error Allowable (4 and 5)) are; pH: 0.02 units on all levels as SD, pC02: 1–2 % (CV) depending on level and the clinical situation. PO,: 2–4%(CV) depending on the level and clinical situation.

Downloads

Download data is not yet available.
Published
1993-09-01
How to Cite
Kofstad J. (1993). 6.1.3.3 Quality Specifications for pH and Blood Gases. Upsala Journal of Medical Sciences, 98(3), 381-385. https://doi.org/10.3109/03009739309179336